Preview

Bulletin of Siberian Medicine

Advanced search

Comparison of cutting and coagulation properties of 1.56 and 1.94 µm fiber lasers and a 0.98 µm semiconductor laser

https://doi.org/10.20538/1682-0363-2021-4-56-62

Abstract

Aim of the study was to compare the cutting and coagulation properties of 1.56 and 1.94 μm fiber lasers with those of a 0.98 μm semiconductor laser.

Materials and methods. A comparative study of the biological effects of 1.56 and 1.94 µm lasers and a 0.98 µm semiconductor laser used in a constant, continuous mode was carried out. The cutting properties of the lasers were evaluated on the chicken muscle tissue samples by the width and depth of the ablation zone formed via a linear laser incision at a speed of 2 mm/s, while the coagulation properties were assessed by the width of the lateral coagulation zone. The zones were measured using a surgical microscope and a calibration slide. For statistical analysis, power values of 3, 5, 7, 9, and 11 W were chosen for each laser wavelength.

Results. Analysis of the findings confirmed that laser wavelength had a statistically significant effect on the linear dependence between incision parameters and laser power. It was found that the 1.56 μm fiber laser (water absorption) had a greater coagulation ability but a comparable cutting ability compared with the 0.98 μm laser (hemoglobin absorption). When used in the power mode of 7W or higher, the 1.94 µm laser provided superior cutting performance compared with the 0.98 µm semiconductor laser at the same exposure power. Elevating the power in any of the lasers primarily increased the width of the ablation zone, and to a lesser extent – the crater depth and the width of the lateral coagulation zone. Therefore, in comparison with the 0.98 μm semiconductor laser, higher radiation power in the 1.56 and 1.94 μm lasers mainly influences their cutting properties, expanding the width and depth of the ablation zone, and has a smaller effect on their coagulation ability.

Conclusion. The findings of the study showed that the 1.56 and 1.94 μm fiber lasers have better coagulation properties in comparison with the 0.98 μm semiconductor laser. was statistically proven that all incision characteristics (width of the lateral coagulation zone, depth and width of the ablation zone) for the 1.56, 1.94, and 0.98 μm lasers depend on the power of laser radiation. The 1.94 µm laser is superior to the 0.98 µm laser in its cutting properties. 

About the Authors

M. A. Ryabova
Pavlov First Saint Petersburg State Medical University
Russian Federation

6-8, L. Tolstogo Str., St. Petersburg, 197022



M. Yu. Ulupov
Pavlov First Saint Petersburg State Medical University
Russian Federation

6-8, L. Tolstogo Str., St. Petersburg, 197022



N. A. Shumilova
Pavlov First Saint Petersburg State Medical University
Russian Federation

6-8, L. Tolstogo Str., St. Petersburg, 197022



G. V. Portnov
Pavlov First Saint Petersburg State Medical University
Russian Federation

6-8, L. Tolstogo Str., St. Petersburg, 197022



E. K. Tikhomirova
Pavlov First Saint Petersburg State Medical University
Russian Federation

6-8, L. Tolstogo Str., St. Petersburg, 197022



M. E. Malkova
Pavlov First Saint Petersburg State Medical University
Russian Federation

6-8, L. Tolstogo Str., St. Petersburg, 197022



References

1. Mooseburger A.S., Herrmann T.R., Liatsikos E., Nagele U., Traxer O. Лазеры и лазерные технологии. Клинические рекомендации Европейской ассоциации урологов; пер. Ю.С. Сиромолот. М.: 2011: 59. URL: https://uroweb.org/wp-content/uploads/21_Lazer.pdf

2. Choy D.S. History of lasers in medicine. Photomed. Laser Surg. 2014; 32(3): 119–120. DOI: 10.1089/pho.2014.9864.

3. Goldman L., Blaney D.J., Kindel D.J. Jr., Franke E.K. Effect of the laser beam on the skin. J. Invest. Dermatol. 1963; 40: 121–122. DOI: 10.1038/jid.1963.21.

4. Тюрин Д.С., Дибиров М.Д., Шиманко А.И., Тебенихин В.С. Арефьев Н., Волков А.С., Цуранов С.В., Швыдко В.С., Магдиев А.Х. Оценка морфологических изменений венозной стенки после эндовазальной лазерной и радиочастотной облитерации. Флебология. 2016; 10 (4): 164–170. DOI: 10.17116/flebo2016104164-170

5. Шайдаков Е.В., Булатов В.Л., Илюхин Е.А., Сонькин И.Н., Григорян А.Г., Гальченко М.И. Оптимальные режимы эндовенозной лазерной облитерации с длиной волны 970, 1 470 и 1560 нм: ретроспективное продольное когортное многоцентровое исследование. Флебология. 2013; 7 (1): 22–29.

6. Янушко В.А., Роговой Н.А., Турлюк Д.В., Климчук И.П., Калинин С.С. Эндовенозная лазерная коагуляция подкожных вен нижних конечностей (1 470 нм и 1 560 нм) в комплексном лечении варикозной болезни. Военная медицина. 2017; 47: 74–78.

7. Шахрай С.В., Гаин М.Ю., Гаин Ю.М., Рябцева С.Н. Экспериментальное обоснование выбора длины волны высокоинтенсивного лазерного излучения для малоинвазивного хирургического лечения хронического геморроя. Казанский медицинский журнал. 2013; 2 (94): 271–277. DOI: 10.17816/KMJ1604.

8. Kang K. et al. The feasibility of 1940-nm diode laser in ton sillectomy. Med. Laser. 2016; 5 (2): 77–82. DOI: 10.25289/ml.2016.5.2.77.

9. Баграташвили В.Н., Баграташвили Н.В., Гапонцев В.П., Махмутоваc Г.Ш., Минаев В.П., Омельченко А.И., Самарцев И.Э., Свиридов А.П., Соболь Э.Н., Цыпина С. И. Изменение оптических свойств гиалинового хряща при нагреве лазерным излучением ближнего ИК диапазона. Квантовая электроника. 2001; 31 (6): 534–538. DOI: 10.1070/QE2001v031n06ABEH001996.

10. Шумилова Н.А. Опыт применения высокоэнергетических лазеров в оториноларингологии. Folia Otorhinolaryngologiae et Pathologiae Respiratoriae. 2016; 22 (1): 75–83.

11. Карпищенко С.А., Рябова М.А., Улупов М.Ю. Лазерная хирургия в оториноларингологии. Consilium Мedicum. 2014; 16 (11): 73–76.

12. Рябова М.А., Улупов М.Ю., Шумилова Н.А. Особенности лазерного воздействия на ткани с длиной волны 1 470 нм. Folia Otorhinolaryngologiae et Pathologiae Respiratoriae. 2016; 22(3): 36–42.

13. Шумилова Н.А., Федотова Ю.С., Рябова М.А. Сравнение биологических эффектов контактного действия полупроводниковых лазеров с длиной волны 1 470 и 810 нм в эксперименте. Современные технологии в медицине. 2014; 6 (4): 62–67.


Review

For citations:


Ryabova M.A., Ulupov M.Yu., Shumilova N.A., Portnov G.V., Tikhomirova E.K., Malkova M.E. Comparison of cutting and coagulation properties of 1.56 and 1.94 µm fiber lasers and a 0.98 µm semiconductor laser. Bulletin of Siberian Medicine. 2021;20(4):56-62. https://doi.org/10.20538/1682-0363-2021-4-56-62